❤️ ECG Biosensor Platform

A low-cost, real-time ECG signal acquisition and analysis platform built from a SparkFun AD8232 and an Arduino Uno R4 Minima — designed, validated, and documented as an independent engineering research project.

Built a $90.78 ECG biosensing platform that achieved 96.3% R-peak detection accuracy and identified a heart-rate-dependent performance limitation through multi-subject validation.

🔗 Live Dashboard: ecg-biosensor.streamlit.app 📄 Published: Oxford Journal of Student Scholarship, Vol. 10, Issue 3 (August 2026) 🎥 Demo Video: Watch on YouTube

ECG Biosensor Dashboard


Project Gallery

System Block Diagram PCB Layout (KiCad, Rev A)
Block Diagram PCB Layout
Raw vs. Filtered Signal Validation Against Apple Watch Series 8
Signal Pipeline Validation Results

Overview

Standard clinical ECG systems cost $1,000–$10,000, placing cardiac monitoring out of reach for many people and resource-limited healthcare settings. This project investigates whether consumer-grade hardware, combined with carefully engineered signal processing, can produce physiologically meaningful cardiac metrics at a fraction of that cost.

Total hardware cost: $90.78

Metric Result
R-peak detection accuracy 96.3% (N=5 trials)
Validated sampling rate 533.3 Hz
Apple Watch agreement (stable conditions) MAPE 3.51–4.10%
Apple Watch agreement (combined, N=25) r=0.762, MAPE=9.30%
Subjects N=6, 30 resting trials

Key Finding

The platform's accuracy is strongly heart-rate-dependent: subjects with resting heart rates above 90 BPM achieved sub-4.5% error against an Apple Watch Series 8 reference, while subjects below 85 BPM saw error rates exceed 16%. This pattern is consistent with a fixed amplitude threshold (median + 0.5 × SD) becoming more susceptible to baseline drift as the interval between true R-peaks grows — and it directly motivates the adaptive threshold work planned for Phase 2.


System Architecture

Electrodes (Ag/AgCl)
        │
        ▼
  AD8232 Analog Front-End  ──►  Arduino Uno R4 Minima (14-bit ADC, 533.3 Hz)
        │                              │
        │                       USB Serial (115200 baud)
        ▼                              ▼
  Right-Leg Drive Circuit       Python Signal Pipeline
                                       │
                         ┌─────────────┼─────────────┐
                         ▼             ▼             ▼
                  Butterworth      IIR Notch     R-peak Detection
                  Bandpass         Filter        (adaptive threshold)
                  (0.5–40 Hz)      (60 Hz)              │
                         └─────────────┼─────────────┘
                                       ▼
                            HRV Metrics (SDNN, RMSSD)
                                       │
                                       ▼
                          Streamlit Real-Time Dashboard

Repository Structure

ecg-biosensor/
├── dashboard/          Streamlit application (live ECG dashboard, CSV playback)
├── docs/               Block diagram, additional documentation
├── experiments/        Protocols for all three structured experiments
├── firmware/           Arduino acquisition code
├── hardware/           KiCad PCB + schematics, Gerber fabrication files
├── report/             Research manuscript and technical report
├── results/            Raw and processed ECG data, all subjects/trials
├── signal_processing/  Filter, peak-detection, and HRV scripts
├── LICENSE
└── README.md

Hardware

Component Part Cost
ECG analog front-end SparkFun AD8232 (SEN-12650) $24.41
Microcontroller Arduino Uno R4 Minima $19.99
Electrodes Ag/AgCl, 100-pack $15.30
Electrode cable CAB-12970 $9.60
Misc. (breadboard, headers, prep pads) $21.48
Total $90.78

A custom 2-layer PCB (140 × 56.5 mm, FR-4, LeadFree HASL) was designed in KiCad following the AD8232 datasheet reference design. Fabrication files were submitted to JLCPCB; boards were received June 13, 2026. Assembly and Phase 2 validation are in progress. See hardware/ for the schematic, PCB layout, and Gerber files.


Signal Processing Pipeline

  1. Bandpass filter — 4th-order Butterworth, 0.5–40 Hz, zero-phase (scipy.signal.filtfilt)
  2. Notch filter — IIR notch at 60 Hz (Q=30) to reject powerline interference
  3. R-peak detection — adaptive amplitude threshold (median + 0.5 × SD), minimum 319-sample inter-peak distance
  4. HRV computation — SDNN and RMSSD from the resulting RR-interval series

See signal_processing/ for the full implementation.


Experiments

# Title Protocol
1 Resting vs. Exercise vs. Recovery 5 resting trials + 3 exercise blocks (Subject 2)
2 Electrode Placement Robustness 4 placement conditions × 3 trials (Subject 2)
3 Filter Algorithm Comparison 3 algorithms × 5 trials (Subject 2)

Full protocols and results are in experiments/ and results/.


Validation

BPM measurements were validated against simultaneous Apple Watch Series 8 recordings across 25 paired readings from five subjects. Apple Watch was selected as an accessible consumer benchmark rather than a clinical gold standard — agreement results should be interpreted as comparative consumer-device performance, not clinical validation.

Subject Avg Apple Watch BPM Mean Bias MAPE
2 97.4 +1.0 3.87%
3 93.8 +1.0 4.10%
4 97.0 +2.6 3.51%
5 72.0 +12.0 16.72%
6 80.2 +14.6 18.28%

Future Work

  • Adaptive, time-decaying threshold envelope to correct the heart-rate-dependent bias
  • PCB assembly and Phase 2 experimental validation
  • Extended multi-subject study (N ≥ 20)
  • Benchmarking against a clinical-grade ECG or Holter monitor
  • Bluetooth wireless data acquisition
  • ML-based arrhythmia classification
  • Battery-powered wearable enclosure

Engineering Design Decisions

Decision Rationale
AD8232 analog front-end Integrated instrumentation amplifier, RLD circuit, and onboard filtering in a single low-cost package ($24.41)
Arduino Uno R4 Minima 14-bit ADC resolution and 48 MHz processor provide higher fidelity than Uno R3; native USB serial reliability
4th-order Butterworth bandpass (0.5–40 Hz) Flat passband preserves P-wave morphology; zero-phase (filtfilt) eliminates timing distortion in batch analysis
IIR notch filter (60 Hz, Q=30) Narrow rejection band removes powerline interference without attenuating adjacent ECG frequencies
Adaptive threshold (median + 0.5 × SD) Scales automatically with signal amplitude across subjects and sessions; more robust than fixed thresholds
533.3 Hz sampling rate Empirically measured (not assumed); exceeds the 250 Hz minimum for reliable QRS detection at 100 BPM
KiCad for PCB design Open-source, industry-standard EDA tool with Python scripting API for automated layout
Streamlit for dashboard Rapid deployment with zero frontend code; cloud-hosted for accessibility without hardware

Development Timeline

Phase 1 (Spring–Summer 2026)
│
├── Hardware assembly — breadboard prototype
│   AD8232 + Arduino Uno R4 Minima + Ag/AgCl electrodes
│
├── Signal processing pipeline
│   Butterworth bandpass → IIR notch → adaptive R-peak detection → HRV
│
├── Experiment 1 — Resting vs. exercise vs. recovery (N=5 trials)
│
├── Experiment 2 — Electrode placement robustness (4 conditions)
│
├── Experiment 3 — Filter algorithm comparison (3 algorithms)
│
├── Multi-subject validation — N=6 subjects, 30 resting trials
│
├── Apple Watch Series 8 validation — N=25 paired readings
│   Key finding: heart-rate-dependent bias identified
│
├── Streamlit dashboard deployed
│   ecg-biosensor.streamlit.app
│
├── KiCad PCB design (Rev A)
│   140 × 56.5 mm, 2-layer FR-4, LeadFree HASL
│
├── Manuscript submitted to Oxford Journal of Student Scholarship
│   Passed editorial screening → entered formal peer review
│
Phase 2 (In Progress)
│
├── PCB boards received from JLCPCB — June 13, 2026
│
└── Assembly and Phase 2 validation — pending

Lessons Learned

Subject 7 exposed a real limitation: Subject 7's resting heart rate fell below 65 BPM, causing near-complete detection failure. Rather than excluding this data point, it was retained as a documented failure case — revealing that the static threshold algorithm has a hard lower bound on reliable performance. This became the central scientific contribution of the paper.

The SNR paradox: Standard SNR metrics penalized the Butterworth filter despite its superior morphological output. The raw signal's dominant DC offset and 60 Hz component inflate Moving Average SNR artificially. This showed that domain-specific evaluation metrics matter more than generic signal metrics for ECG filter quality.

PCB design is iterative: The initial KiCad layout had 52 DRC violations from component overlap and clearance errors. Resolving these iteratively — rather than starting over — taught systematic debugging of hardware design files.

Breadboard instability is a real experimental confound: Motion artifact during post-exercise recording was severe enough to invalidate the filtered pipeline. The custom PCB directly addresses this limitation by eliminating loose wire connections.


Release History

Version Milestone
v0.1 Basic Arduino serial read — raw ECG signal to terminal
v0.2 Moving average filter applied — first visible QRS complex
v0.3 Butterworth bandpass + IIR notch pipeline — clean PQRST morphology
v0.4 Adaptive threshold R-peak detection — 96.3% accuracy
v0.5 HRV computation (SDNN, RMSSD) from RR intervals
v0.6 Streamlit real-time dashboard deployed
v0.7 Multi-subject validation (N=6, 30 trials)
v0.8 Apple Watch Series 8 validation — heart-rate-dependent bias identified
v1.0 Manuscript submitted to Oxford Journal of Student Scholarship
v1.1 Manuscript accepted for publication after peer review
v1.2 Published — Oxford Journal of Student Scholarship, Vol. 10, Issue 3 (August 2026)
v2.0 Custom KiCad PCB (Rev A) — fabricated and received June 2026
v2.1 PCB assembly and Phase 2 validation (in progress)

Custom PCB (Rev A) received from JLCPCB on June 13, 2026.

Bare PCB Shipment Packaging
PCB Packaging

Assembly and Phase 2 validation against the breadboard prototype are in progress.


Running the Dashboard Locally

git clone https://github.com/shrimaan-rapuru/ECG-BIOSENSOR.git
cd ECG-BIOSENSOR/dashboard
pip install -r requirements.txt
streamlit run ecg_dashboard_demo.py

The hosted version at ecg-biosensor.streamlit.app runs in CSV playback mode and requires no hardware.


Ethics Statement

This platform is intended solely for educational and engineering research purposes. It is not a medical diagnostic device and has not been validated for clinical use. All participants were anonymized as Subjects 2–7; no identifying health data is disclosed. Participation was voluntary with informed consent obtained prior to data collection.


Author

Shrimaan Rapuru The Early College at Guilford, Greensboro, NC